Sunday, July 21, 2024

Avoiding baseplate screw failure in reverse total shoulder arthroplasty - innovation without increased cost or complexity


In followup to our post Reverse baseplate failure - fracture of the central screw, the senior author of Avoiding Glenoid Baseplate Fixation Failure by Altering Surgical Technique for Varying Bone Densities offered that "one technical issue leading to incomplete seating is very dense subchondral bone which can provide false sense of the implant being fully seated; another pitiful is if there is some cartilage left on the glenoid which also can lead to incomplete seating booth of these have happened with screw breakage at the interface of the baseplate screw junction"

His conclusions can be summarized as follows:
(1) baseplate failure is an import complication of reverse total shoulder arthroplasty
(2) time zero (immediate post op) stability of the baseplate is important because (a) the patient will load the baseplate before any bone ingrowth has had time to occur and (b) micro motion of the baseplate will inhibit bone ingrowth so that the screws may be all that is holding the baseplate to the glenoid bone leading to the risk of fatigue fracture
(3) failure to remove cartilage and interposed tissue between baseplate and bone may prevent adequate seating of the baseplate
(4) as is the case with all screw fixation in bone, bone quality has a strong effect on the quality of fixation: (a) low bone density may result in lack of a solid "bite" (screw stripping); (b) high bone density may result in incomplete seating
(5) from plain radiographs, the surgeon can preoperatively get an idea of the glenoid bone density: poor (below left), sclerotic (below right).


(6) for bone suspected of having poor bone quality, the author uses some straightforward innovations that do not add time, technology or expense: 
    (a) for soft bone the glenoid is reamed over a 2.5 mm drill (without using a tap) followed by insertion of the baseplate in the untapped bone 
    (b) for denser than average bone,  a 3.0-mm drill is used for the pilot hole, and then the tap is passed (inserted and removed) a total of 3 times, followed by insertion of the baseplate in the tapped bone 


    (c) the standard technique is used for average density bone: a 2.5-mm drill is inserted in the central hole to a depth of 30 mm, followed by use of the 6.5-mm tap, reaming of the glenoid, and finally, insertion of the monoblock baseplate



In in vitro testing using a low-density block model, the standard technique gave a compressive force of 112 N compared with 300 N for the soft bone technique.

In the high-density bone model, the standard technique resulted in failure to seat the baseplate, or screw breakage. Performing the dense bone technique, the baseplate was seated without failure, with an average compressive force of 450 N. 

Comment: Time zero (initial) fixation is important for minimizing the risk of baseplate failure. This requires adequate bone preparation for maximal baseplate-bone contact, full seating of the baseplate and placement of the peripheral screws in good quality bone.

The inspection of preoperative plain radiographs and the use of the techniques suggested here can reduce the risk of inadequate compression on one hand and incomplete seating on the other. 

It is noted that with this implant system, the only screw providing substantial compression is the central one - the others are locking screws. 

Finally, in cases where the central screw is stripped, we have found that "match stick" strips of cortical bone can be progressively added to the central hole until good compression is achieved.

Comments welcome at shoulderarthritis@uw.edu

You can support cutting edge shoulder research that is leading to better care for patients with shoulder problems, click on this link


Follow on twitter/X: https://x.com/RickMatsen
Follow on facebook: https://www.facebook.com/shoulder.arthritis
Follow on LinkedIn: https://www.linkedin.com/in/rick-matsen-88b1a8133/


Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link).
How to x-ray the shoulder (see this link).
The ream and run procedure (see this link).
The total shoulder arthroplasty (see this link).
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
The smooth and move procedure for irreparable rotator cuff tears (see this link).

Shoulder rehabilitation exercises (see this link). 

  

Friday, July 19, 2024

Reverse total shoulder complication - fracture of central baseplate screw

Baseplate failure is a serious complication of reverse total shoulder arthroplasty. See Reverse total shoulder, the complication of baseplate failure.

Here's an example case. An active man had multiple prior attempts at rotator cuff repair, but eventually developed cuff tear arthropathy. 


This was treated with a reverse total shoulder arthroplasty (RSA).


Of note, the baseplate is not in contact with the genoid bone superiorly (see gap at arrow)


The patient returned to physical activity including pushups. A year after the RSA, the shoulder became painful on use. X-rays showed superior tilting of the glenosphere and bending of the central screw.

Eventually the central screw fractured and the glenosphere fixation failed.

Here is an example of fatigue fracture of the central screw.



This case exemplifies the points made by the authors of 
Factors affecting fixation of the glenoid component of a reverse total shoulder prothesis, specifically that the security of baseplate fixation depends in large part on (1) supporting contact between the upper aspect of the baseplate and glenoid bone and (2) strong fixation of the inferior screw in good glenoid bone.



When such support is lacking, there is a risk of fatigue fracture of fixation screws and baseplate loosening.

Loss of superior support for the baseplate can occur when the glenosphere is inferiorly inclined. In other words, inferior inclination requires substantial glenoid reaming to achieve support for the superior baseplate.



Some related articles are referenced below.

How to avoid baseplate failure: the effect of compression and reverse shoulder arthroplasty baseplate design on implant stability


Avoiding Glenoid Baseplate Fixation Failure by Altering Surgical Technique for Varying Bone Densities


Reverse shoulder glenoid loosening: an evaluation of the initial fixation associated with six different reverse shoulder designs

Comment: We invited a very experienced RSA surgeon to give us his perspective. He kindly responded "I too have this failure mode but i think it relates to imperfect seating.  
If there is no bony contact of the baseplate inferiorly that leads to lack of ingrowth. There is a small gap inferiorly too of the baseplate. 
The lack of ingrowth is why the screws failed at the interface to the screw/baseplate junction. I have not had failures with lack of superior contact 
if there is inferior  contact."

Comments welcome at shoulderarthritis@uw.edu

You can support cutting edge shoulder research that is leading to better care for patients with shoulder problems, click on this link


Follow on twitter/X: https://x.com/RickMatsen
Follow on facebook: https://www.facebook.com/shoulder.arthritis
Follow on LinkedIn: https://www.linkedin.com/in/rick-matsen-88b1a8133/


Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link).
How to x-ray the shoulder (see this link).
The ream and run procedure (see this link).
The total shoulder arthroplasty (see this link).
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
The smooth and move procedure for irreparable rotator cuff tears (see this link).

Shoulder rehabilitation exercises (see this link).