Friday, January 26, 2024

Anatomic total shoulder - is a guidewire helpful for reaming?

As pointed out in Edge displacement and deformation of glenoid components in response to eccentric loading. The effect of preparation of the glenoid bone and The radiographic evaluation of keeled and pegged glenoid component insertion, a key to anatomic glenoid component durability is the reaming of the glenoid bone surface so that its concavity exactly fits and completely supports the convex back of the glenoid component as it is exposed to eccentric loading.






Here's an example of poor glenoid preparation.


that can result when a guide wire fixes the orientation of a cannulated reamer in an attempt to modify glenoid version (see How much does "corrective glenoid reaming" actually change version and is the version change of benefit ?)

Poor glenoid preparation cannot be remedied by placing cement between the bone and component - this cement will fracture, displace and leave the component unsupported, often in the critical posterior aspect of the joint.




In contrast to the guide wire and cannulated reamer ("A" below), the wireless approach uses a non-cannulated nubbed reamer ("B" below) that can be angulated so that the bony glenoid face is completely reamed to a single concavity with minimal bone removal. 




While reaming, the surgeon can adjust the angle of the reamer so that blades contact and ream the entire glenoid bone surface (green arrows below).




The adequacy of the reaming can be verified by using a round backed trial to see if it tips when loaded (upper figure) or if it is stable on the reamed glenoid surface (bottom figure).





Wireless reaming can be used with any implant system offering a nubbed reamer that can be inserted into a shallow hole in the center of the glenoid.


    
As described above, the surgeon can adjust the angle of the reamer so that blades contact and prepare the entire glenoid bone surface (green arrows below).


And again, the completeness and adequacy of the reaming can be verified by using a round backed trial to see if it tips when loaded or if it is stable on the reamed glenoid surface.


Below are the preoperative and postoperative axillary truth views from a patient who had wireless reaming for B2 glenohumeral pathology.


The wireless preparation of the glenoid does not require a preoperative CT scan.

It also reduces the risk of over reaming with loss of the glenoid bone stock needed to support the glenoid component as shown in the upper two figures below. (see Version Correction via Eccentric Reaming Compromises Remaining Bone Quality in B2 Glenoids: A Computational Study)








Thanks as always to Steve Lippitt for his wonderful art - often used, seldom acknowledged.

You can support cutting edge shoulder research and education that are leading to better care for patients with shoulder problems, click on this link.

Follow on twitter: https://twitter.com/RickMatsen or https://twitter.com/shoulderarth
Follow on facebook: click on this link
Follow on facebook: https://www.facebook.com/frederick.matsen
Follow on LinkedIn: https://www.linkedin.com/in/rick-matsen-88b1a8133/

Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link).
How to x-ray the shoulder (see this link).
The ream and run procedure (see this link).
The total shoulder arthroplasty (see this link).
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
The smooth and move procedure for irreparable rotator cuff tears (see this link).
Shoulder rehabilitation exercises (see this link).

Sunday, May 15, 2016

3D image intensifier navigation of K-wire placement in the glenoid - statistical vs clinical significance


Improved accuracy of K-wire positioning into the glenoid vault by intraoperative 3D image intensifier-based navigation for the glenoid component in shoulder arthroplasty

These authors hypothesized that navigated k-wire placement is more accurate and shows a smaller deviation angle to the standard centerline compared to the classical “free hand technic” in thirty-four pairs of fresh female sheep scapulae

We include some details of the method:
First, the shoulders from the navigated group 1 were mounted onto the operating table in the upright position. A small incision wasmade in the area of the scapular spine and a carbon-clamp withthree tracking points was fixed. A preoperative scan was performed with Ziehm VisionFD Vario 3D. The middle of the scanner was placed at the central point of the glenoid. The scan performed consisted of 110 single images witha radius of 136◦within 110 seconds. The DICOM-format raw datawere consequently transferred from the 3D image converter to theVectorVision©navigation system. In the navigation system, the data record was used as a CT data record. After creating a 3D image, the data was verified by comparing the image and the anatomy. This control was performedat three defined points (proximal glenoid, central glenoid and the distal glenoid).The instruments were calibrated using the instrument calibration matrix. A 1.8-mm navigable drill sleeve was used. By using the navigation screen, the wire wasplaced as centrally as possible. Subsequently, a K-wire was positioned and shortened to fit into the glenoid. 
In the freehand group 2, the shoulders were stabilized in the upright positionon the operating table. In contrast to group 1, the central point of theglenoid vault was determined visually using electrocauterization. Subsequently, drilling was also performed with the same navigable 1.8 mm sleeve in a perpendicular direction to the glenoid face planeand marked with a K-wire.



 The relation to glenoid standard and alternative centerlines (CL) and the position within the glenoid vault were analyzed.
Results

In groups 1 and 2 the angle between the K-wire and standard CL was 2.2° and 4.7°, respectively (P = 0.01). The angle between the K-wire and alternative CL was 14.4° for group 1 and 17.2° for group 2 (P = 0.02). More navigated K-wire positions were identified within a 5 mm corridor along the glenoid vault CL (52 vs. 39; P = 0.004).

Comment: This article represents yet another method for placing a K-wire in the scapula, ostensibly to guide the preparation of the glenoid for glenoid component placement.

We do not know the cost and time for implementing this approach, whether it would be applicable in the clinical situation or whether the improvement of three degrees would lead to better results for patients than other methods.

As is well know to the readers of this blog, we do not use a guide wire, intaoperative imaging, or patient specific instrumentation to guide the reaming of the glenoid. By contrast (as shown in "D" below), we position a nubbed reamer without a guide wire so that the minimal amount of bone is removed without an attempt to 'correct' glenoid version. Using this method and soft tissue balancing without or with an anteriorly eccentric humeral head component (see this link), we have yet to experience problems with posterior instability or glenoid component failure. In this nice figure from Dr Lippitt, one can see some other approaches and their potential downsides.