Tuesday, August 31, 2021

Dull reamers can kill glenoid bone

Thermal effects of glenoid reaming during shoulder arthroplasty in vivo

These authors point out that glenoid component loosening is a common cause of failure of total shoulder arthroplasty. It has been proposed that the heat generated during glenoid preparation may reach temperatures capable of producing osteonecrosis at the bone-implant interface. They hypothesized that temperatures sufficient to induce thermal necrosis can be produced with routine drilling and reaming during glenoid preparation for shoulder arthroplasty in vivo. Furthermore, they hypothesized that irrigation of the glenoid during reaming can reduce this temperature increase. They used real-time, high-definition, infrared thermal video imaging to determine the temperatures produced by drilling and reaming during glenoid preparation in ten consecutive patients undergoing total shoulder arthroplasty. The maximum temperature and the duration of temperatures greater than the established thresholds for thermal necrosis were documented. 



The first five arthroplasties were performed without irrigation and were compared with the second five arthroplasties, in which continuous bulb irrigation was used during drilling and reaming. A one-dimensional finite element model was developed to estimate the depth of penetration of critical temperatures into the bone of the glenoid on the basis of recorded surface temperatures.

The first hypothesis was supported by the recording of maximum surface temperatures above the 56°C threshold during reaming in four of the five arthroplasties done without irrigation and during drilling in two of the five arthroplasties without irrigation. The estimated depth of penetration of the critical temperature (56°C) to produce instantaneous osteonecrosis was beyond 1 mm (range, 1.97 to 5.12 mm) in four of these patients during reaming and one of these patients during drilling, and two had estimated temperatures above 56°C at 3 mm. 

The second hypothesis was supported by the observation that, in the group receiving irrigation, the temperature exceeded the critical threshold in only one specimen during reaming and in two during drilling. The estimated depth of penetration for the critical temperature (56°C) did not reach a depth of 1 mm in any of these patients (range, 0.07 to 0.19 mm).











They concluded that temperatures sufficient to induce thermal necrosis of glenoid bone can be generated by glenoid preparation in shoulder arthroplasty in vivo. Frequent irrigation may be effective in preventing temperatures from reaching the threshold for bone necrosis during glenoid preparation.


These authors suggest that inaccurate reaming and thermal osteonecrosis from heat generated during the reaming process may contribute to TSA failure by creating a suboptimal bone-implant interface. They investigated the differences in depth penetration and heat generation of used community glenoid reamers in comparison to previous unused reamers.


They used a MTS Servohydraulic machine to test new and used community glenoid

reamers by applying the clinically relevant force of 54.7 N over a defined time. The depth of

penetration was measured via the MTS machine and the thermal profile was obtained via an

infrared camera. The used reamers were then set by the MTS machine to reach the same depth as the new reamers for all respective sizes while recording the force differential generated and capturing the thermal profile. 


At a constant force and time, the new reamers penetrated a greater depth (4.18 mm ± 2.17 mm) than the community used reamers (0.41 mm ± 0.22 mm), a difference of 3.80 mm ((95% CI, 2.23 mm to 5.31 mm), p < 0.001) without generating temperatures above 50°C. 


When programmed to reach the same average depth as the new reamers of equivalent sizes, the community reamers generated more heat on average (50.02 °C ± 2.88 °C), a difference of 5.98 °C ((95% CI, 3.40 to 8.53), p < 0.001). The used reamers on average also required 218.20 N more force than the new reamers (54.71 N ± 28.69 N) to reach the same depth, with the medium (303.47 N ± 96.71 N) and large (261.72 N ± 55.28 N) reamers specifically requiring the largest amount of force.


They concluded that the sharpness of glenoid reamers varies in the community. In order to reach the necessary depth for adequate fixation of implants, orthopedic surgeons may be required to exert a substantially larger force when using dulled reamers in comparison to sharp reamers. As a consequence, the heat generated could increase the risk of thermal osteonecrosis contributing to glenoid loosening.


Comment: These articles point to the ability of reaming to generate bone-killing levels of heat in the glenoid bone. This heat may interfere with the healing of bone in a ream and run procedure and with the stability of fixation in total shoulder arthroplasty.



Follow on twitter: https://twitter.com/shoulderarth

Follow on facebook: https://www.facebook.com/frederick.matsen

Follow on LinkedIn: https://www.linkedin.com/in/rick-matsen-88b1a8133/


How you can support research in shoulder surgery Click on this link.


Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link)
The smooth and move for irreparable cuff tears (see this link)
The total shoulder arthroplasty (see this link).
The ream and run technique is shown in this link.
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
Shoulder rehabilitation exercises (see this link).

Monday, February 22, 2021

Glenoid reaming can generate enough heat to kill bone - relevance to total shoulder and ream and run arthroplasty

Thermal effects of glenoid reaming during shoulder arthroplasty in vivo

These authors point out that glenoid component loosening is a common cause of failure of total shoulder arthroplasty. It has been proposed that the heat generated during glenoid preparation may reach temperatures capable of producing osteonecrosis at the bone-implant interface. They hypothesized that temperatures sufficient to induce thermal necrosis can be produced with routine drilling and reaming during glenoid preparation for shoulder arthroplasty in vivo. Furthermore, they hypothesized that irrigation of the glenoid during reaming can reduce this temperature increase. They used real-time, high-definition, infrared thermal video imaging to determine the temperatures produced by drilling and reaming during glenoid preparation in ten consecutive patients undergoing total shoulder arthroplasty. The maximum temperature and the duration of temperatures greater than the established thresholds for thermal necrosis were documented. The first five arthroplasties were performed without irrigation and were compared with the second five arthroplasties, in which continuous bulb irrigation was used during drilling and reaming. A one-dimensional finite element model was developed to estimate the depth of penetration of critical temperatures into the bone of the glenoid on the basis of recorded surface temperatures.

The first hypothesis was supported by the recording of maximum surface temperatures above the 56°C threshold during reaming in four of the five arthroplasties done without irrigation and during drilling in two of the five arthroplasties without irrigation. The estimated depth of penetration of the critical temperature (56°C) to produce instantaneous osteonecrosis was beyond 1 mm (range, 1.97 to 5.12 mm) in four of these patients during reaming and one of these patients during drilling, and two had estimated temperatures above 56°C at 3 mm. 

The second hypothesis was supported by the observation that, in the group receiving irrigation, the temperature exceeded the critical threshold in only one specimen during reaming and in two during drilling. The estimated depth of penetration for the critical temperature (56°C) did not reach a depth of 1 mm in any of these patients (range, 0.07 to 0.19 mm).

They concluded that temperatures sufficient to induce thermal necrosis of glenoid bone can be generated by glenoid preparation in shoulder arthroplasty in vivo. Frequent irrigation may be effective in preventing temperatures from reaching the threshold for bone necrosis during glenoid preparation.


These authors suggest that inaccurate reaming and thermal osteonecrosis from heat generated during the reaming process may contribute to TSA failure by creating a suboptimal bone-implant interface. They investigated the differences in depth penetration and heat generation of used community glenoid reamers in comparison to previous unused reamers.


They used a MTS Servohydraulic machine to test new and used community glenoid

reamers by applying the clinically relevant force of 54.7 N over a defined time. The depth of

penetration was measured via the MTS machine and the thermal profile was obtained via an

infrared camera. The used reamers were then set by the MTS machine to reach the same depth as the new reamers for all respective sizes while recording the force differential generated and capturing the thermal profile. 


At a constant force and time, the new reamers penetrated a greater depth (4.18 mm ± 2.17 mm) than the community used reamers (0.41 mm ± 0.22 mm), a difference of 3.80 mm ((95% CI, 2.23 mm to 5.31 mm), p < 0.001) without generating temperatures above 50°C. 


When programmed to reach the same average depth as the new reamers of equivalent sizes, the community reamers generated more heat on average (50.02 °C ± 2.88 °C), a difference of 5.98 °C ((95% CI, 3.40 to 8.53), p < 0.001). The used reamers on average also required 218.20 N more force than the new reamers (54.71 N ± 28.69 N) to reach the same depth, with the medium (303.47 N ± 96.71 N) and large (261.72 N ± 55.28 N) reamers specifically requiring the largest amount of force.


They concluded that the sharpness of glenoid reamers varies in the community. In order to reach the necessary depth for adequate fixation of implants, orthopedic surgeons may be required to exert a substantially larger force when using dulled reamers in comparison to sharp reamers. As a consequence, the heat generated could increase the risk of thermal osteonecrosis contributing to glenoid loosening.


Comment: These articles point to the ability of reaming to generate bone-killing levels of heat in the glenoid bone. This heat may interfere with the healing of bone in a ream and run procedure and with the stability of fixation in total shoulder arthroplasty.



Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link).
How to x-ray the shoulder (see this link).
The ream and run procedure (see this link).
The total shoulder arthroplasty (see this link).
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
The smooth and move procedure for irreparable rotator cuff tears (see this link).
Shoulder rehabilitation exercises (see this link).

Thursday, February 23, 2012

Glenoid component fixation. Vacuum assist glenoid fixation: does this technique lead to a more durable glenoid component? JSES

JSES October 2011 published an article Vacuum assist glenoid fixation: does this technique lead to a more durable glenoid component? in which suction is applied to the glenoid vault as a means of removing fluid trapped within the glenoid vault and increasing the amount of cement in the glenoid bone for fixation of a glenoid component. They found that using a special device (that is no longer available) to apply suction to the bone the amount of cement in the glenoid was increased and the number of radiolucent zones were diminished. Of note, however, is the observation that even with this method, the authors had 33% radiolucent lines on postoperative x-rays using the Bigliani-Flatow system.

As we have pointed out in a previous post, more cement can generate more heat in the glenoid bone, risking the possibility of thermal damage to the bone, so more cement may not be the goal.

We use a different approach to the removal of fluid from the prepared glenoid bone, employing a carbondioxide spray just before the cement is pressurized.
The consistent result is the absence of radiolucent lines with a minimal amount of cement as shown in the radiograph below.


--

Use the "Search the Blog" box to the right to find other topics of interest to you.

You may be interested in some of our most visited web pages including: shoulder arthritis, total shoulder, ream and run, reverse total shoulder, CTA arthroplasty, and rotator cuff surgery.

Wednesday, June 29, 2011

Failed, Unsatisfactory Shoulder Joint Arthroplasty - glenoid component loosening

As we saw in the June 27 post 59% of patients coming to us for consultation because they were dissatisfied with their total shoulder joint replacement arthroplasty had loose or worn glenoid components.



The concern about glenoid component failure comes from careful followup of patients having shoulder replacements in the past using a variety of older component designs and older techniques. 

Our shoulder fellow Lazarus published a nice paper pointing to the challenges of achieving a perfect glenoid component insertion and the effect of surgeon experience.

Ideal contouring of the bone beneath the glenoid component (as shown below) is essential as pointed out by shoulder fellow Collins.

When the bone is not properly prepared, there is a tendency to pack in a bit of cement to fill in the defect ('putty carpentry'). This is commonly seen in the posterior (back) half of the glenoid as shown below left. However, this wedge of cement is brittle and not fixed to either the component or bone so it can slip out leaving the component unsupported and at risk for loosening (below right).






From a very well done study of 333 total shoulders, here is a graph showing a 50% radiographic survivorship of glenoid components put in 10 years ago.



Loosening seems to be particularly common when the glenoid design consists of a keel that is fixed into the bone of the shoulder blade using a wad of cement as shown below left and in the color photograph of a retrieved glenoid component




Cement cures with an exothermic reaction - this heats the bone to an amount determined by the amount of cement used. Excessive cement can result in thermal death of the bone with loss of fixation as shown by the thermal camera images below.


Removing a loose glenoid component leaves a big hole in the glenoid bone, making subsequent reconstruction difficult.


We have vigorously investigated the causes of failure and ways in which the security of the component can be optimized. We continue to use a modern glenoid component in the majority of our shoulder joint replacements, because it provides predictable recovery of shoulder comfort and function

Time will tell whether the newer design and our modern techniques will improve the long term durability of the glenoid component.


--

Use the "Search the Blog" box to the right to find other topics of interest to you.

You may be interested in some of our most visited web pages including: shoulder arthritis, total shoulder, ream and run, reverse total shoulder, CTA arthroplasty, and rotator cuff surgery.

Wednesday, May 4, 2011

Total shoulder replacement: avoiding heat damage

One of the concerns about fixation of the glenoid component in total shoulder replacement is the observation that 'radiolucent lines (RLL's)' may develop between the bone on one hand and the component and cement on the other. These RLL's suggest that the connection of the component to bone is not as strong as it should be. Of greater concern is that these RLLs are often progressive, that is, the fixation of the component becomes less secure with time. This is well shown in the article by Walch et al in the SUNDAY, MAY 1, 2011 post. 


There has been little study of the factors that may contribute to the development and progression of RLL's, especially in cases where no lucent lines are seen immediately after surgery.

One factor that we have studied is the possibility of heat damage to the bone of the glenoid causing local death and resorption of the bone over time with progressive mechanical loosening.  In a study with shoulder fellows Churchill, Boorman and Fehringer, Glenoid cementing may generate sufficient heat to endanger the surrounding bone, we found that the amount of heat generated was related to the volume of cement used in fixing the glenoid component. This is because cement cures with an exothermic, or heat generating reaction. With larger amounts of cement, the temperature of bone can rise to a level that can kill the bone.

This observation led us to explore methods of glenoid fixation that used only minimal amounts of cement: precise 'carpentry' with the removal of a minimal amount of bone.

In second study by shoulder fellows Clinton and Lynch along with resident Olson, we found that reaming of the bone could cause thermal damage: Thermal effects of glenoid reaming during shoulder arthroplasty in vivo. The figure below was taken during surgery using a thermal camera. 

We found that irrigating the wound with cool saline solution at the time of reaming kept the temperature of the bone within safe limits.


--

Use the "Search the Blog" box to the right to find other topics of interest to you.

You may be interested in some of our most visited web pages including: shoulder arthritis, total shoulder, ream and run, reverse total shoulder, CTA arthroplasty, and rotator cuff surgery.