How to choose a humeral component in shoulder arthroplasty: ASES Podcast 155


Hat's off to Drs Chalmers and Waterman for yet another outstanding ASES podcast: Episode 155, in which they were joined by Drs Athwal, Cuff and Hatzidakis to discuss fixation of humeral arthroplasty components: "standard" length stems (100-150 mm), short stems (60-90 mm) and stemless designs. 

Here are some takeaways from that presentation coupled with some additional thoughts from the U.W..

(1) The evolution in humeral components is attributed to (a) mitigation of complications (periprosthetic fracture, stress shielding from high canal-filling ratios, early loosening from certain ingrowth surfaces), (b) matching competitor features, and (c) responding to market forces.

The Standard Stem

(2) A standard length stem uses the canal as an alignment guide. It can be stabilized in the canal by (a) a bony ingrowth surface, (b) cement, or (c) reaming to a tight diaphyseal fit, reaming that, in any case, removes cortical bone asymmetrically and unpredictably. [7]. As discussed later in this post, the standard stem can also be stabilized in the canal by impaction grafting. Much of the stress-shielding literature conflates two different things — stem length and canal fill — and they are not the same. Proximal stress shielding tracks with the filling ratio — how much of the canal the implant occupies and how much load it diverts away from proximal bone — far more than with the absolute length of the stem. [9,10]. A long stem that fills the canal shields bone; a long stem that sits loosely in the canal, transferring load through impacted cancellous bone, need not. 

(3) Osteoporosis or altered humeral anatomy may drive the use of a longer stem for better fixation and durability.

The Short Stem

(4) The short stem depends on loading at the metaphyseal level — in bone of variable shape and quality — making both fixation and orientation challenging.

CT of the proximal humerus showing the thin metaphyseal cortex 


The short stem may not have enough leverage to reduce loosening.


Canal filling can lead to stress shielding




And periprosthetic fracture



The absence of the control provided by a standard length stem 
can lead to malpositioning of a short stem




(5) Stemless implants depend on fixation at the anatomic neck, where a thin cortical shell surrounds cancellous bone. Bone density at this level is of variable quality, especially in patients in the arthroplasty age range. While some surgeons claim “100% stemless,” that claim is not realistic for every patient having shoulder arthroplasty. 


AP radiograph with the anatomic-neck fixation zone annotated


The big question about stemless is how a surgeon decides, in a given patient, whether good fixation can be achieved. Preoperative imaging may give some clues, but it comes down to the intraoperative findings. Some have advocated the “thumb” test;


a more reliable approach is to insert the trial nucleus and determine whether it is solid. When it is not solid because of “mushy” cancellous bone, the options are limited. In our experience, adding cancellous graft at the anatomic neck does not rescue fixation: the site is already failing in cancellous bone, and loose graft packed against soft cancellous bone has nothing rigid to lock against. 
The ability to improve fixation with longer fins is likewise limited; upsizing the trial to gain purchase can risk the lateral cortex.
 

It is also worth noting that even when stemless fixation is achieved, adaptive bone changes are common; at short-to-mid term, however, they have not reliably correlated with loosening or worse function.[6,8]

So in the real world, the stemless-loving surgeon needs a Plan “B” when trial fixation is unstable. Conversion to a stemmed component is the most attractive option. One caveat: if the problem is mushy cancellous bone, fixation of a short stem can itself be challenging — a fatter stem versus bone graft, and the same question of how to assure durability in the desired position.

Avoiding Trouble with the Stemless

Be aware of common technical mistakes - assure complete head resection and avoid excessive varus or valgus. 

Insufficient head resection


Varus cut

The free-hand cut is key.

Revision

(6) Two concerns arise when the humeral component needs revision. First, removal of a well-fixed implant can risk fracture of the tuberosity, bone loss, and shaft fracture, and may require a humeral osteotomy or window. Second, although some humeral implants are “convertible” (the humeral fixation system stays in place), retaining the implant only makes sense if it is well fixed at an appropriate height and acceptable version. One podcast participant reported that 25–40% of nominally convertible stems cannot actually be converted in practice. Taken together, the routine use of “convertible” implants seems unattractive.


Malrotation


Too low or too high position of the stem




Another approach

(7) Our experience is that secure, safely revisable humeral component fixation can be achieved with a smooth, standard-length stem set at a small filling ratio and fixed with impaction autografting — using bone from the resected humeral head that one of our fellows named “God’s Own Glue” (see "Procrustean Method") [2,3]. The graft is impacted between a smooth stem and the endosteal cortex, where it stabilizes the implant at the time of surgery, rather than relying on a bony ingrowth surface, cement, or a tight diaphyseal press fit. [4]. A low filling ratio protects the humerus from stress shielding. Because the stem is deliberately undersized, load is transferred through grafted cancellous bone rather than bypassed down a canal-filling implant — addressing stress shielding at the variable that drives it. [9,10]. 

In 48 ream-and-run and 78 total shoulder arthroplasties using a smooth, standard-length impaction-grafted stem, two-year radiographs showed adaptive changes that were generally minor and not associated with component shift or subsidence. Inserted this way, a smooth standard-length stem offers secure, bone-preserving fixation providing results can serve as a basis for comparison for other component designs and fixation methods. [5]. 

In a consecutive single-surgeon series of 458 anatomic total shoulder arthroplasties using this construct (mean follow-up 9.2 years; 114 shoulders beyond ten years), Simple Shoulder Test scores improved from 3.3 to 9.2 and were sustained — never declining by more than the MCID — past ten years. The overall revision rate was 2.6% (12 of 458). None of the revisions were performed for a humeral-component cause: there was no humeral component loosening, no periprosthetic fracture, and no stem-related failure.[11] 




 A low filling ratio protects the humerus from stress shielding

 

Two year followup


Six year followup

Impaction grafting with a low filling ratio avoids incomplete seating of the humeral stem (humerus captivus)


Impaction grafting a standard stem in revising a failed short stem



Impaction grafting a thin long stem  in humeral deformity



Impaction grafting after fracture fixation




Impaction grafting a thin standard stem enables easy, safe removal followed by new implant insertion should revision be necessary.

Thus "revisability" does not require "convertibility"



Returning to the observation stated at the beginning of this post: humeral component evolution is attributed to (a) mitigation of complications, (b) matching competitor features, and (c) responding to market forces. 

Impaction grafting of a thin, smooth stem is an approach for addressing  periprosthetic fracture, stress shielding from high canal-filling ratios, and early loosening. It is not driven by market forces. Impaction grafting can be used with any standard length stem (anatomic or reverse) combined with freely available autograft from the patient’s own humeral head that would otherwise be discarded.


Should we be thinking straight?




Beauty in Simplicity

Tundra Swan
Union Bay Natural Area

Disclosure. The author has no financial relationship with any manufacturer of the orthopaedic devices discussed in this post.



References of interest

1. The ASES Podcast (American Shoulder and Elbow Surgeons), Episode 155: standard, short, and stemless humeral components and convertible designs. Available at: https://www.youtube.com/watch?v=q_s_g8oQgPg

2. Razfar N, Reeves JM, Langohr GDG, Willing R, Athwal GS, Johnson JA. Comparison of proximal humeral bone stresses between stemless, short stem, and standard stem length: a finite element analysis. J Shoulder Elbow Surg. 2016;25(7):1076–1083. doi:10.1016/j.jse.2015.11.011. PMID 26810016.

3. Reeves JM, Langohr GDG, Athwal GS, Johnson JA. The effect of stemless humeral component fixation feature design on bone stress and strain response: a finite element analysis. J Shoulder Elbow Surg. 2018;27(12):2232–2241. doi:10.1016/j.jse.2018.06.002. PMID 30104100.

4. Synnott S, Langohr GDG, Reeves JM, Johnson JA, Athwal GS. The effect of humeral implant thickness and canal fill on interface contact and bone stresses in the proximal humerus. JSES Int. 2021;5(5):881–888. doi:10.1016/j.jseint.2021.05.006.

5. Aibinder WR, Uddin F, Bicknell RT, Krupp R, Scheibel M, Athwal GS. Stress shielding following stemless anatomic total shoulder arthroplasty. Shoulder Elbow. 2023. doi:10.1177/17585732211058804. PMID 36895609.

6. Raiss P, Edwards TB, Deutsch A, Shah A, Bruckner T, Loew M, Boileau P, Walch G. Radiographic changes around humeral components in shoulder arthroplasty. J Bone Joint Surg Am. 2014;96(7):e54. doi:10.2106/JBJS.M.00378. PMID 24695931.

7. Denard PJ, Raiss P, Gobezie R, Edwards TB, Lederman E. Stress shielding of the humerus in press-fit anatomic shoulder arthroplasty: review and recommendations for evaluation. J Shoulder Elbow Surg. 2018;27(6):1139–1147. doi:10.1016/j.jse.2017.12.020. PMID 29422391.

8. Sheth MM, Kahsai EA, Yang J, Whitson AJ, Matsen FA III, Hsu JE. What is the clinical importance of radiographic changes around the humeral component in anatomic shoulder arthroplasty? A minimum 4-year follow-up study. J Shoulder Elbow Surg.2025;34(8):1877–1885. doi:10.1016/j.jse.2024.11.024. PMID 39800107.

9. Denard PJ, Hsu JE, Whitson A, Neradilek MB, Matsen FA III. Radiographic outcomes of impaction-grafted standard-length humeral components in total shoulder and ream-and-run arthroplasty: is stress shielding an issue? J Shoulder Elbow Surg.2019;28(11):2181–2190. doi:10.1016/j.jse.2019.03.016. PMID 31272887.

10. Kim SC, Park JH, Bukhary H, Yoo JC. Humeral stem with low filling ratio reduces stress shielding in primary reverse shoulder arthroplasty. Int Orthop. 2022;46(6):1341–1349. doi:10.1007/s00264-022-05383-4.

11. Lee M, Chebli C, Mounce D, Bertelsen A, Richardson M, Matsen FA III. Intramedullary reaming for press-fit fixation of a humeral component removes cortical bone asymmetrically. J Shoulder Elbow Surg. 2008;17(1):150–155. doi:10.1016/j.jse.2007.03.032. PMID 18029200.

12. Boorman RS, Hacker SA, Lippitt SB, Matsen FA III. A conservative broaching and impaction grafting technique for humeral component placement and fixation in shoulder arthroplasty: the Procrustean method. Tech Shoulder Elbow Surg.2001;2(3):166–175. doi:10.1097/00132589-200109000-00004.

13. Hacker SA, Boorman RS, Lippitt SB, Matsen FA III. Impaction grafting improves the fit of uncemented humeral arthroplasty. J Shoulder Elbow Surg. 2003;12(5):431–435. doi:10.1016/s1058-2746(03)00053-3. PMID 14564262.

14. Lucas RM, Hsu JE, Gee AO, Neradilek MB, Matsen FA III. Impaction autografting: bone-preserving, secure fixation of a standard humeral component. J Shoulder Elbow Surg. 2016;25(11):1787–1794. doi:10.1016/j.jse.2016.03.008. PMID 27262410.

15. Lee et al. Stress shielding effects of short stem alignment and bone density in reverse shoulder arthroplasty. J Orthop Res. 2026. doi:10.1002/jor.70140.

16. Vasiliadis AV, Giovanoulis V, Lepidas N, Bampis I, Servien E, Lustig S, Gunst S. Stress shielding in stemmed reverse shoulder arthroplasty: an updated review. SICOT J. 2024;10:37. doi:10.1051/sicotj/2024029.

17. Ritter D, Raiss P, Denard PJ, Werner BC, Müller PE, Woiczinski M, Wijdicks CA, Bachmaier S. Volumetric humeral canal fill ratio effects primary stability and cortical bone loading in short and standard stem reverse shoulder arthroplasty: a biomechanical and computational study. J Imaging. 2024;10(12):334. doi:10.3390/jimaging10120334.

18. Kramer M, Olach M, Zdravkovic V, Manser M, Raiss P, Jost B, Spross C. The effects of length and width of the stem on proximal humerus stress shielding in uncemented primary reverse total shoulder arthroplasty. Arch Orthop Trauma Surg. 2024. doi:10.1007/s00402-023-05129-w.

19. John PB, Nageswaran S. Mechanobiological evaluation of solid and multiple porous humeral stem architectures in reverse shoulder arthroplasty based on design and materials: a finite element study. Front Bioeng Biotechnol. 2026;13:1675726. doi:10.3389/fbioe.2025.1675726.

20. Takayama K, Ito H. Association between the canal filling ratio and bone resorption in trabecular metal stems in reverse total shoulder arthroplasty: a radiographic analysis using tomosynthesis. JSES Int. 2024;8(5):1077–1086. doi:10.1016/j.jseint.2024.05.010.







Sunday, January 5, 2025

Stemless humeral component - is stress shielding an issue?

Every type and design of humeral component changes the way in which load is transferred from the articular surface to the bone. Thus, adaptive changes (read: "stress shielding) are inevitable.

The authors of Stress shielding following stemless anatomic total shoulder arthroplasty set out to assess the radiographic proximal humeral bone adaptations seen following stemless anatomic total shoulder arthroplasty. They prospectively followed 152 stemless total shoulder arthroplasties utilizing a single implant design 
performed by 21 shoulder arthroplasty surgeons.


At 2 years postoperatively, "stress shielding" was noted in 61 (41%) shoulders. A total of 11 (7%) shoulders demonstrated major stress shielding with 6 occurring along the medial calcar. At the final follow-up, no humeral implants were radiographically loose or migrated. There was no statistically significant difference in clinical and functional outcomes between shoulders with and without stress shielding. Thus the clinical importance of these adaptive changes was not demonstrated.

In reviewing some of the figures from this article, "stress shielding" may not be the principal cause of bone resorption (yellow arrow). Consider this comparison of the 3 month (left) and the 2 year (right) anterior posterior views of the same shoulder. The 2 year film was interpreted as "demonstrating severe stress shielding along the medial calcar".  Note, however that the cement originally beneath the glenoid component (green arrow) is no longer seen on the 2 year film and that the two year film appears to show shift of the glenoid component's position and lucencies around the pegs (red arrow). These findings bring up the possibility of particulate debris as a contributing factor to the humeral bone resorption at the yellow arrow.



Another figure (below) shows sequential axillary radiographs (L to R: 6 weeks, 6 months, 12 months, and 24 months postoperatively) and was interpreted as showing moderate stress shielding along the medial calcar and greater tuberosity regions. However, it appears that that the proximal humeral bone may have been cystic (yellow arrow) leading to what may be humeral component loosening (red arrows).


Thus it is difficult to know whether the changes seen in the humeral bone are a result of "stress shielding" or are a result of problems with humeral and glenoid component fixation, such as those shown below in other figures from this paper.


While the authors described that "Glenoid radiolucent lines were noted in 28 (18%) patients. These were Lazarus grade 1 in 6, grade 2 in 5, grade 3 in 13, and grade 4 in 2", the relationship of these findings to humeral adaptive changes was not reported.

These authors concluded "the occurrence and severity of bony adaptations had no effect on short-term patient clinical outcomes."

For more detail on this topic see Stemless humeral components: stress shielding

You can support cutting edge shoulder research that is leading to better care for patients with shoulder problems, click on this link

Follow on twitter/X: https://x.com/RickMatsen
Follow on facebook: https://www.facebook.com/shoulder.arthritis
Follow on LinkedIn: https://www.linkedin.com/in/rick-matsen-88b1a8133/

Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link).
How to x-ray the shoulder (see this link).
The ream and run procedure (see this link).
The total shoulder arthroplasty (see this link).
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
The smooth and move procedure for irreparable rotator cuff tears (see this link)
Shoulder rehabilitation exercises (see this link).


Saturday, December 28, 2024

Reverse total shoulder: what is the ideal humeral stem?

In contrast to the situation in an anatomic shoulder arthroplasty (below left), the center of rotation (center of yellow circle) in a reverse total shoulder (below right) is not closely aligned with the humeral medullary canal (red line). This results in a rotational moment arm (blue line) and a bending force each of which can contribute to loosening of the humeral stem in the humeral canal.



These loosening factors need be resisted by secure and durable fixation of the prosthesis to bone.

Recently there has been interest in short stemmed reverse humeral components.

However the short stem places the implant at risk for shift in component position as shown below.


Some have tried to manage this tendency for loosening by using tight fitting components, but these can increase the risk of stress shielding as shown below.



And can be difficult to revise in case of malposition, infection or fracture.


I found this recent article co-authored by Patrick Denard of relevance: Volumetric Humeral Canal Fill Ratio Effects Primary Stability and Cortical Bone Loading in Short and Standard Stem Reverse Shoulder Arthroplasty: A Biomechanical and Computational Study. 

The authors used cadaveric specimens to evaluate the effect of three-dimensional volumetric humeral canal fill ratios of reverse shoulder arthroplasty short and standard stems on the biomechanical stability and bone deformations in the proximal humerus. 

They compared standard stems inserted with low and high filling ratios and short stems inserted with low and high filling ratios. 


The specimens were cyclically loaded while optical recording allowed for spatial implant tracking and the quantification of cortical bone deformations in the proximal humerus. 

Both short and standard-length stem RSA humeral components implanted with a low canal filling ratio maintain dynamic bone loading in the medial calcar of the humerus similar to the native bone tested. This dynamic bone loading is desirable in the prevention of bone losss from stress shielding. Thus high canal fill ratios ( > 0.72) may correlate with the severity of bone resorption in the clinical setting.

However, short stems implanted with a low filling ratio (< 0.72) have an increased risk for implant tilt and subsidence compared to those inserted with high filling ratios or standard length stems. Short stems inserted with higher filling ratios and standard length stems implanted with low or high filling ratios demonstrated higher primary stability.

 
A conclusion from the foregoing is that short stems are at risk for stress shielding if they are inserted with high filling ratios, and at risk for loosening (tilt and subsidence) if they are inserted with low filling ratios. 

So, one can reduce the risks of stress shielding and component loosening through the use of a standard length smooth stem inserted with a low filling ratio as seen in the x-ray below with a stable humeral component and no bone resorption two years after reverse total shoulder arthroplasty. Secure fixation is achieved through impaction autografting.

You can support cutting edge shoulder research that is leading to better care for patients with shoulder problems, click on this link

Follow on twitter/X: https://x.com/RickMatsen

Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link).
How to x-ray the shoulder (see this link).
The ream and run procedure (see this link).
The total shoulder arthroplasty (see this link).
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
The smooth and move procedure for irreparable rotator cuff tears (see this link)
Shoulder rehabilitation exercises (see this link).