Showing posts with label navigation. Show all posts
Showing posts with label navigation. Show all posts

Sunday, March 1, 2026

Surgeon and industry views on digital technology


Digital technology in shoulder arthroplasty: what 
do surgeons want? What will industry offer? This well-done survey by Warner and colleagues asked 192 shoulder arthroplasty surgeons from three international organizations (ASES, SECEC, and the Codman Shoulder Society) what digital technologies they use and want, and asked six industry leaders about their development priorities. 

The survey found that 

(1)  Ninety-six percent of respondents perform preoperative planning, 82% use digital tools for this, and about a quarter use navigation or mixed reality intraoperatively. Seventy-six percent of respondents perform more than 40 arthroplasties per year; the study did not reflect the usage or views of the low shoulder arthroplasty volume surgeons who perform the majority of these procedures. The technology adoption question looks very different for a surgeon doing 15 arthroplasties a year versus 100. Industry's implicit pitch is often that technology can compensate for lower surgical volume, but that claim is itself unproven and worth calling out directly.

(2) When asked what they wanted from industry, surgeons prioritized artificial intelligence / machine learning, navigation, and improved virtual reality planning. Industry respondents matched these preferences, ranking AI/ML and navigation as their top development priorities. Both groups felt that AI could reduce errors, generate data for best practices, and improve efficiency.

(3) Surgeons emphasized reliability and patient outcome tracking as priorities, which industry did not consider a key focus. In other words, surgeons want to know if their interventions are working for patients, but industry is more interested in the tools themselves than in measuring whether those tools make a difference.

Significantly, the survey did not ask whether surgeons have observed improved patient-reported outcomes (PROs) when using these technologies: “Has your adoption of digital technology resulted in measurably better patient outcomes?” See Navigation, where are we going?

The Discussion section states that preliminary literature shows improved accuracy of glenoid component positioning with navigation, patient-specific instrumentation, and augmented reality versus freehand techniques. The critical question is whether positioning precision within a narrow range translates into better function, less pain, greater durability, or fewer complications for patients. See: How much precision do we need to pay for in shoulder arthroplasty?

The cost question nobody wants to answer

One of the most revealing findings is that industry respondents could not, or would not, articulate a revenue model for these digital technologies. The paper notes that “no consistent business model for AI/ML emerged” and that companies “were not able, or not willing, to clarify revenue models for most digital technologies.”

This is significant. If companies cannot explain how these tools will be paid for, the cost will inevitably be bundled into implant pricing, passed to hospitals, and ultimately borne by patients and the healthcare system. For technologies without demonstrated clinical benefit, this raises serious questions about value. 

There are other costs of which the surgeon needs to be aware: increased preoperative and OR time, increased space requirements for a robot, time for training in the use of the technology, learning how to detect when the technology is pointing toward a suboptimal procedure, knowing when and how to abort the use of the technology, and the influence the vendor of the technology may have over the choice of implants.

Warner and colleagues have provided a useful snapshot of surgeon and industry attitudes toward digital technology in shoulder arthroplasty. The data on current usage patterns and preferences are informative. They conclude that “further adoption of these technologies will likely be contingent on well done scientific study which demonstrates value.”  Until we have rigorous evidence that these expensive digital tools produce measurably better patient outcomes, we should be cautious about the assumption that more technology equals better care.

The bottom line is that robotics, navigation, and virtual/augmented reality are all methods of transferring a plan made from a preoperative set of images to the patient, without knowing such key characteristics as condition of the soft tissues (that control mobility and stability) and the condition of the bone (that determines quality of fixation).  Such factors loom large in the outcome realized by the patient and often require substantial modifications of the preoperative plan.

Helmuth von Moltke the Elder wrote: “Kein Operationsplan reicht mit einiger Sicherheit über das erste Zusammentreffen mit der feindlichen Hauptmacht hinaus.” i.e., “No battle plan survives contact with the enemy.”

This is not an argument against planning. It’s an argument for: Flexibility over rigidity. Preparation over prediction. Adaptation over perfection. Once real-world complexity hits (fog of war, friction, shoulder variability), reality diverges from the preconceived plan.

The goal isn’t to be precise/accurate to the preoperative plan.
The goal is to be ready to modify it.

See How to make good decisions in shoulder (and other) surgery : Bayesian Thinking - adjusting a prior plan in consideration of new information.

Not infrequently, plans need to be modified,


Cactus Wren
Tucson
2020


Follow on twitter/X: https://x.com/RickMatsen

Follow on facebook: https://www.facebook.com/shoulder.arthritis
Follow on LinkedIn: https://www.linkedin.com/in/rickmatsenmd

Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link).
How to x-ray the shoulder (see this link).
The ream and run procedure (see this link).
The total shoulder arthroplasty (see this link).
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
The smooth and move procedure for irreparable rotator cuff tears (see this link)
Shoulder rehabilitation exercises (see this link).


Thursday, February 12, 2026

Intraoperative navigation for shoulder arthroplasty - where are we going?

Computer-assisted navigation systems for shoulder replacement surgery are - like robotics (see What should be the role of robotics in shoulder arthroplasty?) - commercially advocated to aid surgeons in positioning components to match a preoperative plan based on a static preoperative CT scan.

Opinions vary among expert surgeons regarding which component positions (e.g. glenosphere position, lateralization, inclination; humeral distalization and lateralization) optimize patient reported comfort and function on one hand and minimize the risk of important complications (such as instability/dislocation and acromial/spine stress fractures) on the other. Furthermore the preoperative plan is formed in the absence of important intraoperative findings (such as soft tissue balance and bone characteristics) that are observed after the shoulder is surgically exposed and that often require the surgeon to modify the preoperative plan.

Let's consider a recent systematic review, The Role of Intraoperative Navigation in Reverse Total Shoulder Arthroplasty and its Impact on Clinical Outcomes: which assessed five studies with nearly 2,000 patients having reverses total shoulder carried out using ExactechGPS Navigation.

Three studies directly compared navigated versus conventional non-navigated reverse shoulder replacement. All three showed the same result: no clinically significant differences in functional outcomes.

Short-term clinical and radiologic outcomes of reverse total shoulder arthroplasty with navigation system in the Asian population: a retrospective comparative study found "the use of a navigation system in rTSA showed no significant difference in clinical outcomes and complications compared to conventional implantation."

Early clinical outcomes following navigation-assisted baseplate fixation in reverse total shoulder arthroplasty: a matched cohort study found: "navigated and non-navigated RSAs yielded similar rates of improvement in range of motion and functional outcome scores"

 Intraoperative navigation system use increases accuracy of glenoid component inclination but not functional outcomes in reverse total shoulder arthroplasty: a prospective comparative study. found "There were no statistically significant differences in ROM, PROMs, and satisfaction between patients receiving computer-navigated and standard RSA at a short-term follow-up. "

The review paper (The Role of Intraoperative Navigation in Reverse Total Shoulder Arthroplasty and its Impact on Clinical Outcomes) does not report comparative cost, surgical time, revision rates, or complication rates for patients receiving navigated and non-navigated reverse total shoulders. Furthermore, the article does not address the learning curve for this technology and whether this is a reasonable undertaking for the many surgeons who perform  relatively few reverse total shoulders per year.

Notably absent from all published studies is a comparative cost analysis. The proprietary nature of navigation system pricing—where capital equipment, disposables, software, and per-case fees create opaque total costs—prevents the field from conducting basic cost-effectiveness evaluation. 

There is yet another issue: that of complications related to the navigation system itself, such as problems with fixation of the device to the coracoid. Coracoid fracture represents an uncommonly reported but specific complication of navigation-assisted shoulder arthroplasty, with an overall reported incidence of approximately 0.05% in experienced hands but significantly higher rates (up to 30%) during the learning curve. 

High intraoperative accuracy and low complication rate of computer-assisted navigation of the glenoid in total shoulder arthroplasty analyzed 16,723 navigated shoulder arthroplasty cases reported 9 coracoid fractures.

GPS NAVIGATION SYSTEM ALLOWS THE SURGEON TO PREPARE THE IMPLANT SITE AS PLANNED ON PREOPERATIVE SOFTWARE IN REVERSE SHOULDER ARTHROPLASTY: 3 coracoid fractures in the first 10 cases

THE USE OF NAVIGATION IN REVERSE SHOULDER ARTHROPLASTY: PRELIMINARY REPORT OF 11 CASES: 1 fracture in the first case.

Safety and Efficacy of Intraoperative Computer-Navigated Versus Non-Navigated Shoulder Arthroplasty at a Tertiary Referral: 2 cases (5.5%) of coracoid fracture.

If you are curious, here's some detail about the ExactechGP Navigation system.  You might like by looking at this video

The Basic Technology -  

The Components:

1. Preoperative Planning

    * Patient gets a CT scan of their shoulder

    * Surgeon uploads CT to Equinoxe Planning App software

    * Surgeon creates a 3D virtual plan: glenoid component position, version, inclination, depth, screw trajectories

    * Within days, the plan is uploaded to GPS workstation

2. Intraoperative Hardware

GPS Workstation:

    * Compact mobile unit within sterile field

    * Has an infrared camera that tracks the LED markers

    * Screen displays real-time surgical guidance

Three Active Trackers (reusable, battery-powered with IR LEDs):

    * Probe Tracker: Hand-held pointer for registration

    * Glenoid Tracker: Attached to coracoid process (bone stabilizer)

    * Tool Tracker: Attached to surgical instruments (reamers, drill guides)

3. The Workflow

Step 1: Registration (Mapping CT to Patient)

    * Surgeon attaches Glenoid Tracker to coracoid process with a pin

    * Using the Probe Tracker, surgeon touches specific anatomic landmarks on the scapula

    * System matches these physical points to the CT scan

    * Creates a "coordinate system" linking the patient's real anatomy to the virtual plan

Step 2: Real-Time Guidance

    * Surgeon attaches Tool Tracker to reamer/drill/guide instruments

    * Camera tracks all three trackers simultaneously via their IR LEDs

    * Screen shows:

        o Current instrument position vs. planned position

        o Version and inclination in real-time

        o Reaming depth

        o Screw trajectories and lengths

Step 3: Verification

    * After component placement, system verifies final position

    * Compares achieved vs. planned placement


Conclusion:

The question each surgeon must ask is whether intraoperative navigation addresses problems experienced by patients in their own practice having reverse total shoulder arthroplasty.

The question the field of shoulder surgery must ask is which specific patient problems (instability? fracture? stiffness? pain?) occur due to positioning errors that: (a) surgeons using standard techniques cannot adequately control, and (b) navigation systems would prevent. 


Finding the way


Snow Geese 
Skagit, Washington
202

Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link).
How to x-ray the shoulder (see this link).
The ream and run procedure (see this link)
The total shoulder arthroplasty (see this link)
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
The smooth and move procedure for irreparable rotator cuff tears (see this link)
Shoulder rehabilitation exercises (see this link). 


Friday, September 15, 2023

Is computer-assisted intraoperative navigation helpful for the patient having reverse shoulder arthroplasty?

The goal of intraoperative navigation is to surgically implement a preoperative computer-generated plan.



The value of this innovation depends on several factors, including (1) whether the plan, if executed, would improve the clinical outcome for patients and (2) how much navigation adds to the time and cost of the procedure.

Many studies of navigation evaluate how close the actual baseplate position is to the preoperative plan (accuracy) and the level of consistency or reproducibility of the implant position (precision). Most do not evaluate whether executing the preoperative plan is likely to improve the patients' clinical outcome.

For example The Value of Computer-Assisted Navigation for Glenoid Baseplate Implantation in Reverse Shoulder Arthroplasty A Systematic Review and Meta-Analysis found that a common goal of preoperative planing is to change preoperative retroversion to a postoperative value of 0 degrees.



To determine whether this is, in fact, a goal that is likely to lead to better outcomes for the patient we can look to Baseplate retroversion does not affect postoperative outcomes after reverse shoulder arthroplasty, a study in which the authors sought to preserve glenoid bone stock without aiming for a certain degree of retroversion. Conservative reaming was only carried out only to the extent necessary to create a single concavity to support the baseplate. They compared the two-year outcomes for 161 patients having postoperative component retroversion ≤ 15 degrees to those for 110 patients with postoperative retroversion >15 degrees. There was less than a 5 degree preoperative to postoperative change in version for each group.

There were no significant differences between the groups in American Shoulder and Elbow Surgeons, visual analog scale, or Single Assessment Numeric Evaluation scores. There were also no significant differences in postoperative range of motion or revision rates.

Thus this study does not provide evidence supporting implementation of a preoperative plan to "correct" preoperative version to 0 degrees. Furthermore, these authors point out that there are complications associated with correction of native retroversion and a lack of consensus on optimal baseplate positioning. 

A recent study Intraoperative navigation system use increases accuracy of glenoid component inclination but not functional outcomes in reverse total shoulder arthroplasty: a prospective comparative study evaluated the clinical utility of intraoperative navigation in patients undergoing RSA. 16 patients had navigation and 17 had standard RSA evaluated at a mean follow-up of 16 months. In both groups, planned and postoperative versions were not significantly different.  Patient reported outcomes, range of motion and satisfaction did not differ between the navigated and non-navigated patients at final follow-up.

The authors of The Value of Computer-Assisted Navigation for Glenoid Baseplate Implantation in Reverse Shoulder Arthroplasty A Systematic Review and Meta-Analysis found essentially no clinical followup data in the reviewed studies. They concluded that computer-assisted navigation surgery seemed to increase the accuracy and precision of glenoid baseplate inclination in achieving the preoperatively planned position. However, they stated that the clinical value of this technology in terms of improving prosthesis longevity, complications, and patient functional outcomes is unknown.

Interestingly, augmented glenoid components were used in 74% of the cases with navigation in comparison to 32% in cases without navigation. The odds ratio of selecting an augmented glenoid component with the use of planning and navigation was 8.1.

The utilization of preoperative planning, intraoperative navigation, augmented glenoid components, and associated increased operating room time added cost in comparison to 2D planning and standard implants. 

Two-year clinical outcomes and complication rates in anatomic and reverse shoulder arthroplasty implanted with Exactech GPS© intraoperative navigation found that for 533 reverse total shoulder arthroplasty using 73% augmented glenoids, the 31-month outcomes were similar to those in a concurrent series of cases in which the surgeon did not elect to use intraoperative navigation and used augmented glenoids in only 26%.

The Impact of Preoperative Three-Dimensional Planning and Intraoperative Navigation of Shoulder Arthroplasty on Implant Selection and Operative Time: A Single Surgeon’s Experience  assessed the impact of preoperative 3D planning on the surgeon’s selection of the glenoid component (standard versus augmented) and compare duration of surgery with and without intraoperative navigation in 200 patients who underwent shoulder arthroplasty. The first group of 100 patients underwent shoulder arthroplasty utilizing standard 2D preoperative planning based on standard radiographs and CT scans. The second group of 100 patients underwent shoulder arthroplasty utilizing 3D preoperative planning and intraoperative navigation. 

For the group of patients with standard preoperative planning, only 15 augmented glenoid components were utilized while in the group of patients with 3D preoperative planning and navigation 54 augments were used.
The operative time was 11 minutes longer for the procedures that used intraoperative navigation, compared to those that did not.

The authors stated, "Since 3D planning and intraoperative navigation is more costly than 2D planning, and augmented glenoid components are more costly than standard glenoid components, the cost-benefit of these changes with respect mid-term and long-term clinical outcomes and implant survival has not been ascertained."

Comment: The value of navigation in executing a preoperative plan depends on whether the plan when executed is likely to improve the outcome for the patient. At present, there is a lack of consensus on the best approach to positioning the glenoid baseplate; for example, should glenoid retroversion be "corrected"? (see Glenoid version: acceptors or correctors). The approach used in Baseplate retroversion does not affect postoperative outcomes after reverse shoulder arthroplasty prioritizes bone preservation rather than changing version and can be realized without intraoperative navigation.

Is computer-assisted intraoperative navigation helpful for the patient having reverse shoulder arthroplasty?

You can support cutting edge shoulder research that is leading to better care for patients with shoulder problems, click on this link.

Follow on twitter: https://twitter.com/shoulderarth
Follow on facebook: click on this link
Follow on facebook: https://www.facebook.com/frederick.matsen
Follow on LinkedIn: https://www.linkedin.com/in/rick-matsen-88b1a8133/

Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link).
How to x-ray the shoulder (see this link).
The ream and run procedure (see this link).
The total shoulder arthroplasty (see this link).
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
The smooth and move procedure for irreparable rotator cuff tears (see this link).
Shoulder rehabilitation exercises (see this link).

Saturday, June 24, 2023

Intraoperative computer-assisted navigation in shoulder arthroplasty.


Intraoperative computer-assisted navigation is a technology designed to help surgeons apply a preoperative plan to the patient in the operating room. One such system is the ExactechGPS (see this video), which utilizes a stereotaxic tracker attached to the coracoid process and identifies the location of pre-selected points on the glenoid and scapula as references for orienting instruments in the performance of the glenoid arthroplasty. 
 







The authors of Two-year clinical outcomes and complication rates in anatomic and reverse shoulder arthroplasty implanted with Exactech GPS© intraoperative navigation  present their results using this system in anatomic (ATSA) and reverse (RTSA) total shoulder arthroplasties performed using preoperative planning and intraoperative navigation.

35 different surgeons contributed cases to this series; the decision to use navigation was left up to the surgeon. 

For anatomic arthroplasty 216 shoulders (65 years of age, 50% female, 11 degrees glenoid retroversion, 57% augmented glenoids) had average 31 month outcomes:155 degrees active flexion, SST score 11, Constant score 75, ASES score 87, UCLA score 32, SPADI score 16, SAS score 83.  Complications included 1 humeral fracture, 5 glenoid loosening, 1 cuff tear, and 2 unexplained pain. 4 (1.9%) were revised.  These outcomes were similar to those in a concurrent series of cases in which the surgeon did not elect to use intraoperative navigation (i.e. the outcomes did not differ by the minimal clinically important difference (MCID)).

For reverse total shoulder arthroplasty 533 shoulders (72 years of age, 56% female, 9 degrees glenoid retroversion, 73% augmented glenoids) 31 month outcomes: 142 degrees active flexion, SST score 10. Constant score 75, ASES score 84, UCLA score 31, SPADI score 22, SAS score 77 Complications included 7 humeral fractures, 3 implant breakage , 1 acromial fracture, 1 coracoid fracture and 1 unexplained pain. 5 (0.9%) were revised.  These outcomes were similar to those in a concurrent series of cases in which the surgeon did not elect to use intraoperative navigation (i.e. the outcomes did not differ by the minimal clinically important difference (MCID)).


For the cases in which navigation was used, surgical time was increased by 8-10 minutes.


Comment: This article presents good clinical outcomes for anatomic and reverse total shoulder arthroplasties using intraoperative navigation with minimal complications associated with this technology. 

It would be of interest to know the authors' indications for the use of navigation (is it being recommended for use in all arthroplasty cases or only cases with challenging pathoanatomy?).  Of note, during the time period of this study, at least twice as many arthroplasties were performed by the authors without navigation as with navigation. 

The value of a technological innovation can be measured as the benefit to the patient divided by its costs. Navigation involves substantial additional instrumentation, training and time in addition what is required for non-navigated arthroplasty; these costs are not presented in this publication.

Additional clinical research is needed to determine the degree to which the incremental costs of intraoperative navigation are offset by better patient outcomes.

You can support cutting edge shoulder research that is leading to better care for patients with shoulder problems, click on this link.

Follow on twitter: https://twitter.com/shoulderarth
Follow on facebook: click on this link
Follow on facebook: https://www.facebook.com/frederick.matsen
Follow on LinkedIn: https://www.linkedin.com/in/rick-matsen-88b1a8133/

Here are some videos that are of shoulder interest
Shoulder arthritis - what you need to know (see this link).
How to x-ray the shoulder (see this link).
The ream and run procedure (see this link).
The total shoulder arthroplasty (see this link).
The cuff tear arthropathy arthroplasty (see this link).
The reverse total shoulder arthroplasty (see this link).
The smooth and move procedure for irreparable rotator cuff tears (see this link).
Shoulder rehabilitation exercises (see this link).


Monday, June 29, 2020

Is Preoperative Three-Dimensional Planning and Intraoperative Navigation of Shoulder Arthroplasty of Value?

The Impact of Preoperative Three-Dimensional Planning and Intraoperative Navigation of Shoulder Arthroplasty on Implant Selection and Operative Time: A Single Surgeon’s Experience

These authors assessed the impact of preoperative 3D planning on the surgeon’s selection of the glenoid component (standard versus augmented) and compare duration of surgery with and without intraoperative navigation in 200 patients who underwent shoulder arthroplasty. 
The first group of 100 patients underwent shoulder arthroplasty utilizing standard 2D preoperative planning based on standard radiographs and CT scans. 
The second group of 100 patients underwent shoulder arthroplasty utilizing 3D preoperative planning and intraoperative navigation. 
For the group of patients with standard preoperative planning, only 15 augmented glenoid components were utilized while in the group of patients with 3D preoperative planning and navigation 54 augments were used. 
The operative time was 11 minutes longer for the procedures that used intraoperative navigation, compared to those that did not.

Comment: While this study showed that the surgeon's choice of implant was affected by the use of preoperative three-dimensional planning and intraoperative navigation of shoulder arthroplasty, it does not show that this effect benefitted the patient in that clinical outcome and survivorship data were not presented. As the authors state, "Since 3D planning and  intraoperative navigation is more costly than 2D planning, and augmented glenoid components are more costly than standard glenoid components, the cost-benefit of these changes with  respect mid-term and long-term clinical outcomes and implant survival has not been ascertained."

Data from the Australian Orthopaedic Association registry, shows that the 10 year revision rate for cross-linked all-polyethylene components inserted without preoperative three-dimensional planning, intraoperative navigation, or augmented glenoid components is below 5% (see below). One must ask "How likely is it that these more expensive technologies will lead to better outcomes than these?"








===


To see a YouTube on how we do total shoulder arthroplasty with an emphasis on glenoid preparation, click on this link.


===



To see our new series of youtube videos on important shoulder surgeries and how they are done, click here.


Use the "Search" box to the right to find other topics of interest to you.


You may be interested in some of our most visited web pages  arthritis, total shoulder, ream and run, reverse total shoulder, CTA arthroplasty, and rotator cuff surgery as well as the 'ream and run essentials'

Thursday, December 11, 2014

What is the value of intraoperative navigation for glenoid component positioning in total shoulder arthroplasty?

Benefit of intraoperative navigation on glenoid component positioning during total shoulder arthroplasty.

These authors conducted a systematic, online search for articles on shoulder arthroplasty using computer guided navigation. They found 6 groups of 247 shoulders from 5 studies that met their inclusion criteria.

The pooled weighted mean difference for deviation from neutral version was -6.4° (95 %CI -7.9 to -5.3) in favor of navigation, but there was no significant difference in tilt.

They concluded that navigation allows for more accurate glenoid version, but the clinical meaningfulness of the absolute improvement over standard techniques is questionable.

Comment: Three issues are surfaced here. (1) While it is recognized that abnormal glenoid version in an arthritic shoulder is associated with inferior clinical results, it has yet to be demonstrated that 'correction' of version to some theoretical 'normal' improves the clinical outcome over the results of arthroplasty that accepts the pathological version and, thereby, preserves glenoid bone stock. Thus, computer guided navigation may lead the surgeon to change the version without consideration of whether this change will be helpful or harmful. (2) There are no data whether an 'improvement' of version of 6.4 degrees has any clinical benefit. (3) There are no data on the cost and time necessary to achieve an 'improvement' of 6.4 degrees in glenoid version. 

So at the end of the day we do not know the value (benefit/cost) of computer guided navigation.

Parenthetically, it is of interest that most of the included studies were on cadavers or computer models as opposed to studies of real surgery on real patients. 

===



To see the topics covered in this Blog, click here

Use the "Search" box to the right to find other topics of interest to you.

You may be interested in some of our most visited web pages including:shoulder arthritis, total shoulder, ream and run, reverse total shoulder, CTA arthroplasty, and rotator cuff surgery as well as the 'ream and run essentials'




Wednesday, July 10, 2013

Computer guidance for a reverse total shoulder, is it of value?

3D navigated implantation of the glenoid component in reversed shoulder arthroplasty. Feasibility and results in an anatomic study


There is a fascination with the use of "3-D computer-guided navigation" to improve the results of shoulder arthroplasty. This study compared 15 navigated and 12 non-navigated cadaveric insertion of the guidewire for glenoid baseplate implantation in reverse shoulder arthroplasty. A Kirschner wire was placed freehand or with the use of a navigated drill guide. For the navigated procedures, a flat detector 3D C-arm with navigation system was used. The goal was to insert the Kirschner wire 12 mm from the inferior glenoid, with an inferior tilt of 10° and centrally in the axial scapular axis.

For the 3D C-arm navigated procedure, the reference array was positioned onto the scapular spine with two 3.0 mm Schanz screws pointing caudally over two small skin incisions. An optoelectronic camera was positioned at a location that enables tracking of both the flat detector 3D C-arm and reference array. Before performing a 3D scan, two-dimensional fluoroscopic images were obtained to identify the correct positioning of the 3D C-arm. For optimum image quality, a carbon table and 3D C-arm system were recommended.

The authors found the inferior glenoid drilling distance was 14.1 ± 3.4 mm for conventional placement and 15.1 ± 3.4 mm for the navigated procedure and that the inferior tilt was 7.4 ± 5.2° for conventional and 7.7 ± 4.9° for the navigated insertion. The glenoid version in the axial plane showed significantly higher accuracy for the navigated procedure, with a mean deviation of 1.6 ±4.5° for the navigated procedure compared with 11.5 ± 6.5° for the conventional procedure.

The authors do not quantitate the increased time and cost of computer guidance. These tests were not carried out in shoulders with the type of pathology encountered in patients needing reverse total shoulders, which may require special consideration of the ideal glenosphere positioning. They confined their analysis to the positioning of the guide wire and did not examine the effectiveness of navigation in placement of the glenosphere positioning. Insertion of two Schanz screws in the scapular spine is a concern and may predispose the spine to fracture. Finally, the radiation exposure is much higher in the navigation approach.

So, all in all, the case is not compelling for this navigation approach.  Instead, we have been impressed with the need to customize the procedure to the special anatomy of each patient - an approach in which navigation to some pre-conceived orientation is not applicable.

===
To see the topics covered in this Blog, click here


Use the "Search" box to the right to find other topics of interest to you.

You may be interested in some of our most visited web pages including:shoulder arthritis, total shoulder, ream and run, reverse total shoulder, CTA arthroplasty, and rotator cuff surgery as well as the 'ream and run essentials'



See from which cities our patients come.




See the countries from which our readers come on this post.

Saturday, February 9, 2013

CT navigation for reverse total shoulder

Intraoperative CT navigation for glenoid component fixation in reverse shoulder arthroplasty

Patients having a reverse total shoulder often represent the nexus of (1) altered shoulder anatomy, (2) poor bone quality, (3) increased loading of the arthroplasty and (4) increased fall risk. For these reasons, fixation of the glenoid component must be optimized. 

In our practice we find that good glenoid exposure allows us to optimize component and screw placement while minimizing the removal of bone and properly orienting the glenoid component, as shown in this recent case on the shoulder of a gentleman from southern California.



The authors of this article present a process that includes the extra steps of placing a positioning pin with infrared markers for intraoperative navigation. Registration was done using an intra operative CT scan. Data were fed to a "navigation suite to facilitate glenoid component fixation". CT images in three planes were used to assess the position of the central peg before reaming. CT images were also used to achieve the best desired screw position.

We agree with the authors that optimizing positioning and screw placement while minimizing the risk of scapular fractures is important, we are yet to be convinced that the cost and increased OR time associated with CT navigation is justified in terms of improved patient outcomes.

Our preoperative planning is based on preoperative high quality plain films and our surgical technique is based on good glenoid exposure.


==

Use the "Search" box to the right to find other topics of interest to you.

You may be interested in some of our most visited web pages including:shoulder arthritis, total shoulder, ream and run, reverse total shoulder, CTA arthroplasty,  and rotator cuff surgery.