Scapular notching in reverse shoulder arthroplasty - it is important!



Anytime we have unintended contact between high density polyethylene and bone, it is a problem. Scapular notching is a radiographic finding, but the real concerns are about (1) the damage to the poly of the humeral cup, (2) loss of the bone of the scapula that supports the glenoid component, and (3) the potential for instability resulting from leverage of one against the other. See this previous post which discusses this phenomenon in some detail.



In the Grammont-type reverse total shoulder, contact of the adducted humeral component against the scapula is not uncommon as shown in this figure from a manufacturer's website.




These authors retrospectively reviewed 448 patients who underwent a Grammont-type reverse total shoulder  (461 shoulders) performed for rotator cuff tear arthropathy or osteoarthritis with cuff deficiency with a mean followup of 51 months (range, 24-206 months). They found notching of the scapula in 68% of the cases; it was present in 48% at one year after surgery. 

Notching was more common in active patients, in patients with cuff tear arthropathy, and in patients with greater degrees of superior displacement of the humeral head before surgery. Strength and range of motion were compromised in patients with notching.

Importantly, 36% of shoulders with notching had humeral radiolucent lines (in contrast to 17% in those without notching), suggesting the possibility that polyethylene particles from the humeral cup causing bone resorption. Similarly glenoid loosening was three times more common in the presence of notching.

The authors point out that standardized plain x-rays are necessary for the evaluation of notching, noting that sometimes notching is better seen on the axillary view.

Comment: Scapular notching is important and can be expected to adversely affect the long term durability and function of the reverse. It is best avoided by (1) use of a glenoid component design that offsets the center of rotation from the scapula, (2) proper positioning of the glenoid component at the inferior aspect of the glenoid, (3) avoiding superior tilt of the glenoid component, and carefully checking for contact between the humeral component and scapula at surgery when the arm is adducted and rotated. If contact is noted after component implantation, careful resection of the contacting scapular bone may be helpful.



===
Consultation for those who live a distance away from Seattle.

Check out the new Shoulder Arthritis Book - click here.

Click here to see the new Rotator Cuff Book

To see the topics covered in this Blog, click here

Use the "Search" box to the right to find other topics of interest to you.

You may be interested in some of our most visited web pages including:shoulder arthritis, total shoulder, ream and runreverse total shoulderCTA arthroplasty, and rotator cuff surgery as well as the 'ream and run essentials'