Left pair: standard-length humeral stems. Right pair: a convertible-platform stem in an anatomic total shoulder (A) and after conversion to a reverse with the stem left in place (B); from Burrus MT, Bedi A, Werner BC. J Am Acad Orthop Surg. 2025;33(7):346-51, Figure 2 [22].
Summary
Orthopaedic companies are in the business of developing and selling components for shoulder arthroplasty. These implants remain on the market for varying lengths of time. Some are recalled. Others are quietly replaced by newer models. Many were never shown to be better than the ones they replaced. Here are some relevant observations:
· New implants are not tested in patients before launch. New shoulder implants reach the market through the FDA's 510(k) "clearance" pathway. It requires only "substantial equivalence" to a device already on the market, usually without clinical data to show increased benefit to patients in comparison to its predecessors. All 476 shoulder arthroplasty devices “cleared” from 1976 to 2021 came to market this way [2]. By contrast, devices such as heart valves must go through Premarket Approval (PMA), which requires clinical evidence of safety and effectiveness. Any new drug (including “minor” modifications of a prior drug) needs controlled clinical trials before it can be sold.
· New implants fail and are recalled. 510(k) “cleared” devices were recalled far more often than devices approved with clinical trials: 17.8% versus 1.6% from 2002 to 2012 [1]. Of 476 shoulder devices cleared since 1976, 80 (16.8%) have been recalled [2].
· New implants are often “cleared” based on previous devices that failed. More than half of the shoulder devices cleared by the FDA in 2020–21 trace their “equivalence”, through a chain of predicates, to a device that was later withdrawn from the market because of design flaws [2].
· Failures of new implants are recognized late. Implant problems are commonly recognized years, sometimes more than a decade, after many of them have been implanted in patients. For the most serious (Class I) orthopaedic recalls from 2002 to 2023, arthroplasty implants were recalled a median of 5.7 years after the first adverse event report reached the FDA [14]. Recalls of total elbow and radial head implants have occurred 4 to 16 years after launch [4].
· New designs displace established ones without evidence that they are of increased benefit to patients. Two decades of new shoulder technologies have not been shown to improve patient outcomes [5, 6].
Before adopting a “new” implant, patients and surgeons may want to inquire how it differs from designs with an established track record, what evidence supports its advantages, and how it was approved for use. Jeanne Lenzer’s book The Danger Within Us shows what can be at stake when these questions go unasked [7].
The story
A patient who had a successful shoulder replacement on one side five years ago may find that their implant is not available for treating their arthritis on the opposite side. This may or may not be because something better came along. Understanding why implants fall out of use begins with understanding how they get into use in the first place.
1. “Clearance” is not proof of value to patients
The FDA has two main routes for implants. Premarket Approval (PMA) requires clinical trials showing safety and effectiveness. The 510(k) pathway (the one usually used for shoulder implants) requires only that the new device be “substantially equivalent” to a device previously cleared by the FDA for sale, its “predicate”. Most 510(k) devices are “cleared” without clinical data showing safety or superiority. One might ask, if the new device is “substantially equivalent” to a previous one, what is its value?
Day and colleagues examined FDA records from 1992 to 2012 [1]. Outside orthopaedics, the share of devices “cleared” through 510(k) fell from 91% to 53%. In orthopaedics it barely moved, from 94% to 88%. Shoulder implants are the extreme case. Zhu and colleagues found that every one of the 476 shoulder arthroplasty devices “cleared” from 1976 to July 2021 came through the 510(k) mechanism. None went through Premarket Approval [2]. The 21st Century Cures Act of 2016 further relaxed some of the FDA’s premarket evidence requirements [16].
2. Implants cleared without trials are recalled more often
Who decides on a recall? Nearly all device recalls are started voluntarily by the manufacturer, in coordination with the FDA; the FDA then assigns each recall a risk class, from Class I (most serious) to Class III [14]. The FDA can also ask a company to recall a device and, when a device could cause serious harm or death, can order a recall, but such orders are rare. From 2002 to 2012, 17.8% of 510(k)-cleared devices were recalled, compared with 1.6% of PMA-approved devices. Among the twenty companies with the most recalls, 510(k)-cleared devices had 11.5 times the odds of recall of PMA-approved devices [1].
For shoulders specifically, Zhu and colleagues reviewed all 476 shoulder arthroplasty devices "cleared" through 510(k) from 1976 to July 2021 [2]. Eighty (16.8%) have since been recalled. The reasons were labeling (19), process control (13), packaging (10), non-implant design issues (10), implant design (10) and other causes (18). The ten implant-design recalls involved five complete shoulder systems, four humeral components and one glenoid component. One was a Class I recall, the FDA’s most serious category, defined as one in which continued exposure can cause serious harm or death [2].
3. Problems from prior devices can be inherited
Because each 510(k) device is “cleared” by pointing to a predicate, shoulder implants form a family tree. Zhu and colleagues mapped it [2]. Two-thirds of all “cleared” shoulder devices have served as a predicate for later ones. Two early designs, the Neer II Total Shoulder (K895226) and the DePuy Global Total Shoulder (K914695), each have more than 300 descendants.
· 130 of the 476 devices (27.3%) are linked to at least one predicate that was later recalled for a design problem.
· Of the 29 shoulder devices “cleared” in the single year from July 2020 to July 2021, 16 (55.2%) were connected, through their chains of predicates, with devices later withdrawn from the market because of design flaws.
· The Zimmer Trabecular Metal Reverse Shoulder (K052906) was itself one of the ten devices recalled for implant design. It was also the most influential of them: 110 later devices trace their “clearance” back to it.
· The Biomet Comprehensive Reverse Shoulder (K080642), cleared in 2008, was the Class I recall. It was withdrawn in August 2017 for higher than anticipated fracture rates due to its design, and had served as predicate for 67 devices [2]. Its humeral tray had already been changed from titanium to cobalt-chromium in 2012 to improve fatigue strength, several years before the Class I recall for fractures of the titanium version [14].

Here is the full figure from Zhu, showing the ancestry in detail.

Redrawn from Zhu and colleagues, Figure 4 [2]. Each dot is a shoulder device cleared through 510(k) whose “equivalence” traces back to the Biomet Comprehensive Reverse Shoulder (K080642), cleared in 2008 and recalled in 2017. Rows show how many generations each device is removed from the Biomet device; the shaded area marks devices cleared after its recall. Zhu A, Ying X, Carey EG, Pean CA, Premkumar A, Fu MC, Taylor SA. J Shoulder Elbow Surg. 2023;32(3):671-6.
The authors note that devices “cleared” decades apart remain linked by “equivalence” despite substantive changes in materials and design. “Substantial equivalence” is not “equivalence” [2].
4. Failures are recognized late
The FDA learns of problems largely through voluntary reports to its MAUDE database. This database is open to surgeons as well but rarely consulted by them.
Reports to MAUDE are incomplete and slow to accumulate. Yet any report of implant failure contains information of use to surgeons. Our analysis of 4,063 shoulder arthroplasty complications reported from 2012 to 2016 shows how much information sits in this database [3]. For three total elbow and radial head implants, recall came 4 to 16 years after market introduction and clinical implantation, even though failure reports had already reached the FDA much earlier [4].
Gedik and colleagues measured this delay directly for the highest-risk orthopaedic recalls [14]. From November 2002 to December 2023, the FDA issued 16 Class I orthopaedic recall events, involving 20 devices; 8 of the events involved arthroplasty implants, 7 spine devices and 1 other. Only 13% of the recalled devices had gone through PMA; the rest were “cleared” through 510(k). Eleven of the 16 recall events (69%) were due to design flaws. For arthroplasty implants, the median time from the first relevant MAUDE report to recall was 5.7 years (interquartile range 3.0 to 7.8), against 1.5 years for spine devices (P = 0.014). Arthroplasty implants had been on the market a median of 8.1 years when they were recalled. Across all 16 recall events, a median of 534 units of the recalled device had already been distributed (interquartile range 82 to 5,626).
The authors suggest that, with litigation costs in the billions, manufacturers may investigate at length before volunteering a recall. They also observe that many of the recalls for faulty design involved a modified version of an existing implant. A manufacturer can change a “cleared” device, for example its material or how its parts connect, and have the new version “cleared” through another 510(k) by showing that it is “substantially equivalent” to the old one, usually without clinical data. The FDA has the authority to order a manufacturer to study how a device performs once it is in patients (a “Section 522” postmarket surveillance study), but it has rarely done so for orthopaedic implants; metal-on-metal hips are an exception. The authors call for closer monitoring when a modification substantially changes how a device works [14].
The DePuy ASR hip shows the scale this can reach. It was recalled in August 2010, seven years after its 2003 European launch, after more than 93,000 had been sold. At the time of recall, UK registry data showed that 12% to 13% were revised at five years; by 2011 the British Orthopaedic Association reported revision rates for the ASR XL of 21% at four years and 49% at six years [8]. By the time an implant falls out of use for poor performance, many patients already have it in them.
5. New is not necessarily better
Many implants are not recalled. They are simply replaced by the next model, sometimes quietly. Gedik and colleagues point out that nearly all recalls are voluntary, so a device with problems can be modified or withdrawn without any formal recall; they cite the DePuy Attune knee baseplate, withdrawn and modified in 2017 after reports of early loosening [14]. Whether the next model is better is rarely tested.
The pace of change is rapid. In a single year, from July 2020 to July 2021, the FDA cleared 29 shoulder arthroplasty devices through 510(k), 18 of them complete implant systems [2]. Zhu and colleagues describe the market as fast-paced and competitive, with companies seeking quick clearance to enter it [2]. Manufacturers have business reasons to keep introducing new products: a new design can be priced and marketed as an advance. But each new technology carries added costs for research, development, FDA clearance, marketing, surgeon advocacy and surgeon training [6]. A BMJ investigation of the ASR hip described companies that “scarcely let a year go by” without a “new improved” implant, a “fashion trade” that adds cost without proven benefit [8]. None of this means new implants are bad, but it means their introduction is driven at least as much by business as by evidence that patients will be better off.
The “convertible” humeral stem is an example. These “platform” stems are designed so that an anatomic total shoulder can later be converted to a reverse without removing the stem. In one study of 102 conversions, those in which the stem could be retained had shorter operations, less blood loss and fewer complications than those in which it was exchanged. Yet the stem had to be removed in 73 of the 102 (72%); the reasons were not given, but malposition or loosening of the stem are likely causes [21]. A 2025 review by Burrus and colleagues, generally favorable to these systems, likewise reports that convertible stems were removed in up to half of revisions, often because they are malpositioned [22]. An alternative to a “convertible” humeral stem is a smooth stem fixed with impaction grafting, like the one shown in the case below, that can be removed easily if conversion is ever needed, so that a reverse stem can be placed at the right height and version [20].
Burrus and colleagues document other limitations. Convertible metal-backed glenoid components have had loosening rates of 20% to 40%; in the 2023 Australian joint registry report, modular metal-backed glenoids had a 10-year revision rate of 20.1%, compared with 7.6% for all-polyethylene glenoids. Furthermore, a convertible implant can be converted only if the company still makes the matching parts, possibly decades after the original surgery [22].
Crosby and colleagues stated that platform stems cost about the same as other modular stems but presented no cost data [21], and we could find no published comparison of what convertible and standard stems cost.
Somerson and colleagues analyzed shoulder arthroplasty studies published from 1990 to 2015 [5]. Adjusted clinical outcomes for anatomic total shoulders were only marginally better in more recent reports (p = 0.048), and revision rates were not. For reverse shoulders, neither outcomes nor revision rates improved.
Schiffman and colleagues reviewed 114 articles on anatomic total shoulder arthroplasty [6]. They looked at preoperative CT, 3-D planning, patient-specific instrumentation, stemless and short-stem humeral components, and metal-backed, hybrid and augmented glenoids. They found no evidence that results improved over two decades, or that any of these technologies was associated with better patient outcomes.
When a new implant is not shown to be better, its adoption may owe as much to marketing as to evidence.
A case: an implant that worked well, but is now discontinued

Left shoulder of a 65-year-old with cuff-intact arthritis: before surgery (left); immediately after total shoulder arthroplasty with an impaction-grafted Global Advantage stem (center); and six years later, with 12 of 12 Simple Shoulder Test functions (right). Note absence of stress shielding or adaptive changes.
A 65-year-old had arthritis of the left shoulder with an intact rotator cuff. The shoulder was treated with a total shoulder arthroplasty using the Global Advantage system. The smooth, canal-preserving, standard-length humeral stem was fixed with impaction autografting, which allowed a stem with a low filling ratio rather than one that fills the humeral canal. Denard and colleagues found this method of fixation secure and bone-preserving at two years in 78 total shoulders and 48 ream-and-run arthroplasties [19]. Six years after surgery, this patient had excellent comfort and function, able to perform all 12 functions of the Simple Shoulder Test.
This humeral stem is no longer available. It did not fall out of use because it failed. It was replaced by newer designs, which have not been shown to give patients better results [5, 6]. In fact, this was my go-to implant for decades. None of the implants available today match its ease of use, bone conservation and low cost, and none surpass its clinical effectiveness for patients.
What to ask before choosing a new implant
· How does this implant differ from designs with long, well-documented track records? Consider bearing surfaces, geometry, materials, fixation and modularity.
· What adverse outcomes could those differences cause?
· Was it cleared through 510(k) or approved through PMA? What were its predicates, and have any been recalled?
· What clinical evidence, not just mechanical testing, shows that it helps patients?
· What do joint registries show about its survival?
· What information about this implant is in the MAUDE database?
Patients can ask their surgeon the same questions, and should keep a record of the exact implant they receive.
Policy
In 2011 the Institute of Medicine recommended replacing 510(k) with a framework giving reasonable assurance of safety and effectiveness throughout a device’s life cycle [17]. A 2012 bill to tighten the pathway, the Safety of Untested and New Devices Act, never received a floor vote [18]. A stricter definition of “substantial equivalence,” and closer scrutiny of devices that are not in fact equivalent, would mean fewer implants falling out of use after patients have already received them. For more on how the FDA regulates shoulder implants, see our earlier post [15].
Epilogue: does the FDA protect our patients and us?

The Danger Within Us, by Jeanne Lenzer (Little, Brown, 2017)
Jeanne Lenzer’s The Danger Within Us: America’s Untested, Unregulated Medical Device Industry and One Man’s Battle to Survive It (Little, Brown, 2017; 336 pages) tells the story of Dennis Fegan [7, 9]. He was a former oil-rig worker and firefighter with severe epilepsy who had a Cyberonics vagus nerve stimulator implanted to control his seizures. About six years later, at age 48, he collapsed repeatedly at home. In the emergency department, the monitor showed his heart stopping for 30 seconds every three minutes, in time with the device’s electrical pulses. His neurologist came to the ER and switched the device off, and his heart rhythm returned to normal [10]. The device had also failed to control his seizures and left a wire and a blood clot in his jugular vein [11].
The stimulator came through PMA, the stricter pathway. Its 2005 expansion to depression was approved only after the head of the FDA’s device center overruled his own reviewers, who were concerned about low response rates and suicide attempts in the trials [12, 13]. If problems like these can slip through PMA, the 510(k) pathway, which requires no clinical data, offers even less protection.
Lenzer, a former emergency-medicine physician associate and longtime BMJ contributor, argues that high-risk devices are tested far less rigorously than drugs, that their harms are poorly tracked, and that financial ties shape how they are used [7, 9]. Marcia Angell, former editor-in-chief of the New England Journal of Medicine, called the book “disturbing and riveting.” It is worth reading for anyone who chooses implants.
Keeping a sharp eye on what’s important

Peregrine Falcon
Follow on twitter: https://twitter.com/shoulderarth
Follow on facebook: https://www.facebook.com/frederick.matsen
Follow on LinkedIn: https://www.linkedin.com/in/rick-matsen-88b1a8133/
References
1. Day CS, Park DJ, Rozenshteyn FS, Owusu-Sarpong N, Gonzalez A. Analysis of FDA-approved orthopaedic devices and their recalls. J Bone Joint Surg Am. 2016;98(6):517-24. PMID 26984921. doi:10.2106/JBJS.15.00286
2. Zhu A, Ying X, Carey EG, Pean CA, Premkumar A, Fu MC, Taylor SA. Shoulder arthroplasty device clearance: an ancestral network analysis. J Shoulder Elbow Surg. 2023;32(3):671-6. PMID 36279987. doi:10.1016/j.jse.2022.09.017
3. Somerson JS, Hsu JE, Neradilek MB, Matsen FA 3rd. Analysis of 4063 complications of shoulder arthroplasty reported to the US Food and Drug Administration from 2012 to 2016. J Shoulder Elbow Surg. 2018;27(11):1978-86. PMID 29759905. doi:10.1016/j.jse.2018.03.025
4. Somerson JS, Matsen FA 3rd. Timely recognition of total elbow and radial head arthroplasty adverse events: an analysis of reports to the US Food and Drug Administration. J Shoulder Elbow Surg. 2019;28(3):510-9. PMID 30466818. doi:10.1016/j.jse.2018.08.043
5. Somerson JS, Neradilek MB, Hsu JE, Service BC, Gee AO, Matsen FA 3rd. Is there evidence that the outcomes of primary anatomic and reverse shoulder arthroplasty are getting better? Int Orthop. 2017;41(6):1235-44. PMID 28353051. doi:10.1007/s00264-017-3443-0
6. Schiffman CJ, Prabhakar P, Hsu JE, Shaffer ML, Miljacic L, Matsen FA 3rd. Assessing the value to the patient of new technologies in anatomic total shoulder arthroplasty. J Bone Joint Surg Am. 2021;103(9):761-70. PMID 33587515. doi:10.2106/JBJS.20.01853
7. Lenzer J. The Danger Within Us: America’s Untested, Unregulated Medical Device Industry and One Man’s Battle to Survive It. New York: Little, Brown; 2017. ISBN 978-0-316-34376-3. Publisher page: https://www.hachettebookgroup.com/titles/jeanne-lenzer/the-danger-within-us/9780316343787/
8. Cohen D. Out of joint: the story of the ASR. BMJ. 2011;342:d2905. PMID 21572134. doi:10.1136/bmj.d2905
9. Publishers Weekly. Review of The Danger Within Us. November 27, 2017. https://www.publishersweekly.com/9780316343763
10. Brownlee S, Lenzer J. Medical devices can kill. Reader’s Digest, September 2010. Reprinted by the National Center for Health Research: https://www.center4research.org/medical-devices-can-kill/
11. Cart LE. Review of The Danger Within Us. Shelf Awareness, January 23, 2018. https://shelf-awareness.com/readers/2018-01-23/the_danger_within_us:_america_s_untested_unregulated_medical_device_industry_and_one_man_s_battle_to_survive_it.html
12. Vagus nerve stimulation device approved with multiple cautions. Psychiatric News. 2005;40(16). doi:10.1176/pn.40.16.00400014a
13. War of the Wires. Mother Jones, November 2005. https://www.motherjones.com/politics/2005/11/war-wires/
14. Gedik CC, Walker C 2nd, Bouloussa H, Krumme JW, Cheng AL, Cil A, Dubin JR. Characterization and time-delay in issuing high-risk recalls of orthopaedic devices from 2002 to 2023. J Am Acad Orthop Surg. 2026;34(12):e1628-e1635. PMID 41453025. doi:10.5435/JAAOS-D-25-00947
15. Related post: How effective is the U.S. Food and Drug Administration (FDA) in assuring safety and effectiveness of shoulder arthroplasty implants? UW Shoulder and Elbow Academy, October 23, 2022. https://shoulderarthritis.blogspot.com/2022/10/how-effective-is-us-food-and-drug.html
16. Kesselheim AS, Avorn J. New “21st Century Cures” legislation: speed and ease vs science. JAMA. 2017;317(6):581-2. PMID 28056124. doi:10.1001/jama.2016.20640
17. Institute of Medicine. Medical Devices and the Public’s Health: The FDA 510(k) Clearance Process at 35 Years. Washington, DC: National Academies Press; 2011. doi:10.17226/13150
18. Safety of Untested and New Devices Act of 2012, H.R. 3847, 112th Congress (2012). https://www.govtrack.us/congress/bills/112/hr3847
19. Denard PJ, Hsu JE, Whitson A, Neradilek MB, Matsen FA 3rd. Radiographic outcomes of impaction-grafted standard-length humeral components in total shoulder and ream-and-run arthroplasty: is stress shielding an issue? J Shoulder Elbow Surg. 2019;28(11):2181-90. PMID 31272887. doi:10.1016/j.jse.2019.03.016
20. Matsen FA 3rd. Humeral stems – when should I buy a convertible? UW Shoulder and Elbow Academy, May 2017. https://shoulderarthritis.blogspot.com/2017/05/humeral-stems-when-should-i-buy.html
21. Crosby LA, Wright TW, Yu S, Zuckerman JD. Conversion to reverse total shoulder arthroplasty with and without humeral stem retention: the role of a convertible-platform stem. J Bone Joint Surg Am. 2017;99(9):736-42. PMID 28463917. doi:10.2106/JBJS.16.00683
22. Burrus MT, Bedi A, Werner BC. Convertible humeral and glenoid components for anatomic shoulder arthroplasty. J Am Acad Orthop Surg. 2025;33(7):346-51. PMID 39937952. doi:10.5435/JAAOS-D-23-01134
23. Australian Government Department of Health, Disability and Ageing. Prescribed List of Medical Devices and Human Tissue Products, effective 1 November 2026. https://www.health.gov.au/resources/publications/prescribed-list-of-medical-devices-and-human-tissue-products