Short stemmed humeral components - is there a case to be made?

Radiographic changes differ between two different short press-fit humeral stem designs in total shoulder arthroplasty

These authors compared the radiographic changes of the humerus in the short term after total shoulder arthroplasty using two different short-stem humeral components,
(1) a collarless humeral stem with an oval geometry and curved stem (Ascend or Ascend Flex;Wright Medical, Memphis, TN, USA) (42 patients) 

and
(2)  a humeral stem with a metaphyseal collar, rectangular geometry, and straight stem (Apex; Arthrex, Inc., Naples, FL, USA) (35 patients).



At a minimum of 2 years postoperatively the authors determined that there were greater radiographic changes seen with the Ascend (classified as low in 38% and high in 62%, medial calcar osteolysis in 71% than with the 
Apex (classified as low in 77% and high in 23%, medial calcar  osteolysis in 28.5%).

The examples they present show a stress shielding effect with both designs.



Comment: This study, along with others highlighted in prior posts, point out that humeri with short stems often demonstrate a stress shielding effect. This is in contrast to the findings with smooth, impaction grafted standard stems (see below) that appear to preserve at least as much bone as a short stemmed prosthesis.


Secure fixation may also become an issue with short stemmed components as shown by this case we saw last week. Here is the x-ray soon after surgery 


and here is the x-ray one year after surgery, showing humeral component loosening.


Here is another recent post on the topic of short stems.

This is the implanted used in that study:



Sixty-four patients with a mean age of 64.1 years were available for follow-up at a mean of 25 months postoperatively.  On radiographic examination, 9% of stems were deemed "at risk" for loosening (defined as the presence of any stem subsidence, varus or valgus tilt, or radiolucent lines ≥2 mm in 3 or more radiographic zones), but there was no gross loosening in any patient. 

They reviewed a recent report (see this link) evaluating the clinical and radiographic outcomes of 73 TSAs performed with a collarless press-fit short stem which demonstrated a 71% rate of radiolucent lines. The overall revision rate at a mean of 33 months was 12%, and 8.7% of the remaining stems were considered at risk for loosening. They reviewed another recent report (see this link) that reported radiolucencies in 22% of 49 short-stem TSAs with a press-fit collared design.

Comment: Short stemmed humeral components are usually described as 'bone preserving". We agree that bone preservation is important, but find that this objective can be achieved with a standard humeral component inserted with impaction grafting - an approach that is at least as bone preserving as with the use of short stemmed implants. Because impaction grafting distributes interface forces over a broad area, a standard length stem without ingrowth surfaces minimizes problems with stress shielding and achieves additional stability through greater bone-prosthesis contact.

Compare the amount of bone preserved with this method (seen below) in comparison to that with the short stemmed component (shown above).

 



Additional thoughts on humeral components can be found here.

=====
The reader may also be interested in these posts:



Consultation for those who live a distance away from Seattle.

Click here to see the new Shoulder Arthritis Book.

Click here to see the new Rotator Cuff Book

Information about shoulder exercises can be found at this link.

Use the "Search" box to the right to find other topics of interest to you.

You may be interested in some of our most visited web pages including:shoulder arthritis, total shoulder, ream and runreverse total shoulderCTA arthroplasty, and rotator cuff surgery as well as the 'ream and run essentials'

See from which cities our patients come.

See the countries from which our readers come on this post.